Logo
User: Guest  Login
Authors:
Datzmann, Gerd; Dollinger, Günther; Goeden, Christian; Hauptner, Andreas; Körner, Hans-Joachim; Reichart, Patrick; Schmelmer, Oliver 
Document type:
Zeitschriftenartikel / Journal Article 
Title:
The Munich microprobe SNAKE 
Subtitle:
First results using 20 MeV protons and 90 MeV sulfur ions 
Title of conference publication:
7th International Conference on Nuclear Microprobe Technology and Applications 
Journal:
Nuclear Instruments and Methods in Physics Research Section B 
Volume:
181 
Issue:
1-4 
Conference title:
International Conference on Nuclear Microprobe Technology and Applications (7., 2000, Bordeaux) 
Venue:
Bordeaux, France 
Year of conference:
2000 
Date of conference beginning:
10.09.2000 
Date of conference ending:
15.09.2000 
Year:
2001 
Pages from - to:
20-26 
Language:
Englisch 
Keywords:
Nuclear microprobe ; Ion beam analysis 
Abstract:
The scanning ion microprobe called Superconducting Nanoscope for Applied nuclear (Kern) physics Experiments (SNAKE) is taken into operation at the Munich 15 MV tandem accelerator. During the first experiments 16 and 20 MeV protons as well as 90 MeV 32S ions were used to test all equipments. With a reduced divergence of the beam, an overall lateral resolution of 700 nm was achieved by scanning a gold grid with a focused 90 MeV sulfur beam and detecting transmitted ions. However, some field distor...    »
 
ISSN:
0168-583X 
Department:
Fakultät für Luft- und Raumfahrttechnik 
Institute:
LRT 2 - Institut für angewandte Physik und Messtechnik 
Chair:
Dollinger, Günther 
Open Access yes or no?:
Nein / No