Abstract:
The neutron-induced positron source Munich (NEPOMUC) provides a monochromatic low-energy positron beam with an intensity of >1×109e+ s-1 and a full width at half maximum (FWHM) diameter of about 10 mm. To create a small beam focus or sharp positron pulses of 100 ps FWHM the beam brightness needs to be enhanced by remoderation. This is achieved by focusing the primary beam magnetically onto a tungsten single crystal W(100) in reflection geometry. Afterwards the beam exhibits an intensity of >5×107e+ s-1 and a diameter of <2 mm FWHM. To further optimize the beam quality of NEPOMUC we redesigned the remoderation unit. The new setup allows a precise positioning of the remoderator crystal within the focus of the magnetic lens. Additionally, a replacement of the crystal within several minutes and without breaking the beamline vacuum is possible that offers the opportunity for systematic tests of different remoderator materials. 